ISA Certified Control Systems Technician (CCST) program
Certified Control System Technicians (CCSTs) calibrate, document, troubleshoot, and repair/replace instrumentation for systems that measure and control level, temperature, pressure, flow, and other process variables.
CCST question
An electronic d/p cell and pressure transmitter device has been suspected of reading incorrectly during some high-pressure plant operations. The instrument supervisor sends her best technician during the next shutdown to diagnose the problem. The as-found condition of the pressure transmitter as recorded by the instrument technician has given the following values:
![]() |
What should the instrument technician report to the supervisor?
A. There is no apparent problem with the pressure sensor/transmitter.
B. The pressure sensor/transmitter is giving a non-linear response; replacement of the sensor and/or the transmitter will be required.
C. The pressure transmitter span has drifted; the pressure transmitter should be recalibrated.
D. The pressure transmitter zero has shifted; the pressure transmitter should be recalibrated.
CCST answer
The correct answer is B, the pressure sensor/transmitter is giving a non-linear response; replacement of the sensor and/or the transmitter will be required.
Response A (no apparent problem) would be appropriate only if the % Deviations calculated for the as-found conditions were all very close to zero (e.g., <0.01%).
Response C (span drift) would be indicated by % Deviations all increasing in the positive direction with a very small deviation (e.g., <0.01%) at the 0% value. In addition, the % Deviation for increasing and decreasing input values would need to be very close in value, indicating hysteresis is not a problem.
Response D (zero shift) would be indicated by % Deviations of roughly all the same magnitude, either (+) or (-), including the 0% values. In addition, the % Deviation for increasing and decreasing input values would need to be very close in value, indicating that hysteresis is not a problem.
From the table of as-found values, the % Deviations cross from positive to negative several times, indicating a non-linear response. In addition, the as-found % Deviation values as the input signal is decreased are quite different from the values found with increasing input signal, indicating hysteresis in the output signal. This may indicate the d/p cell is damaged or the pressure transmitter electronics are not functioning properly.
Reference: L.D. Goettsche; Maintenance of Instruments & Systems, 2nd Edition, ISA Press, 2003.
Read questions answered by our experts or join the email list.


